Built on the proven technology of the Handicare C-Series, the new A-450 and A-625 ceiling lifts raise patient handling to the next level. Increase patient protection with antimicrobial materials. Simplify patient transfers with its tilting design. Easily access real-time data and diagnostics. Handicare’s A-Series ceiling lifts give you the advantage in safe patient handling.
Unique antimicrobial strap provides reduced infection risks, enhanced durability and a visual indicator for replacement when needed
Emergency safeguard features include: auto shut-off with audible low battery indicator, emergency stop and power lowering
Available with manual or power traversing and constant charge and return to charge options
Manual Traverse | Power Traverse | Power Traverse RTC | Power XY
Lift Motor
24 VDC
Traverse Motor
24 VDC (Optional at time of Purchase)
XY Traverse Motor
Charger
Model: Soneil, 2403SRM20 Input: 100-240 VAC 1.5 Amps, 50-60 Hz Output: 24 VDC, 1.0 Amps Note: Only use the Soneil, 2403SRM30 or Mascot 9940 chargers as alternates with the C-450/C625 Lift. Other chargers, including the A450/A625 charger, are not compatible to the C-450/C-625.
Batteries
24 VDC (2 x 12 VDC) 5.0 AH, Sealed Lead Acid, UR/UL
Lift Case
Flame Retardant ABS
Hand control
Pneumatic Main Unit IP Rating IP22 IP22 = Protected against access to hazardous part by finger, protected against vertically falling water drops up to 15 degrees
Lifting Range
Up to 96” (2438mm)
Lift Weight
21—23.5 lb (9.5-10.65 kg)
Maximum Safe Working Load:
450 lb / 204.5 kg (C-450) or 625 lb / 284 kg (C-625). Note: The maximum safe working load (SWL) of the installed lift is determined by referring to the product label located on the bottom of the lift.
Duty Cycle
1 Min On—9 Mins Off
Max Sound Level
Raising Max load 56.1 dB, Lowering Max load 60.3 dB on a lift in good working order.
Service Life
22,500 cycles or 10 years, whichever comes earlier. As a precautionary measure, the lifting strap should be replaced every 5000 cycles or 3 years, whichever comes earlier. Higher usage lifts may necessitate more frequent replacement of the lifting strap. See “General Inspection and Preventive Maintenance.” Usage Environmental
Conditions
Temperature: +5º to +40º C (41 to 104 °F) Relative Humidity: 15 to 90% RH Atmospheric Pressure: 70 to 1060 hPa
Storage Environmental Conditions
Temperature: -15º to +40º C (5 to 104 °F) Relative Humidity: 10 to 100% RH Atmospheric Pressure: 500 to 1060 hPa
Rated Performance
30-40 lifts at 625 lb (284 kg) 50-60 lifts at 450 lb (204.5 kg) 1 min “on”, 9 min “off”36 Each lift being 24” (610 mm) at the middle of the lifting range (from 54” (1370 mm) strap out to 30” (762 mm) strap out) per full battery) Note: The lift has a break in period; breaking in of the lift will need to be done before these numbers will be achieved. The breaking in period varies from lift to lift and depends on the frequency of use and the types of load being applied. Higher loads and a greater frequency of use break in the lift faster.